WTW 202706 Panduan Pengguna Kit Perawatan Probe DO Tugas Berat

For CellOx Probes

1. Pendahuluan

This manual provides comprehensive instructions for the proper use and maintenance of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit. This kit is designed to ensure the optimal performance and longevity of your WTW CellOx Dissolved Oxygen (DO) probes.

The kit includes essential components for routine care, such as cleaning solution, abrasive foil, membrane heads, and electrolyte solution, all crucial for accurate and reliable dissolved oxygen measurements.

2. Isi Paket

The WTW 202706 Heavy-Duty DO Probe Maintenance Kit contains the following items:

  • Cleaning solution for oxygen sensors (30 ml)
  • Electrolyte solution for oxygen sensors (50 ml)
  • Abrasive foil
  • Three membrane heads
Contents of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit, including two bottles of solution, a clear protective cap, a black probe component, and a yellow protective tube.

Gambar 1: Lebihview of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit components. Visible are bottles of cleaning and electrolyte solutions, a protective cap, a probe component, and a yellow protective tube.

3. Pengaturan dan Persiapan

Before performing maintenance, ensure you have a clean, dry workspace and any necessary personal protective equipment (PPE).

  1. Identify Probe Type: Confirm that your DO probe is a WTW CellOx probe, as this kit is specifically designed for these models.
  2. Gather Components: Lay out all components from the maintenance kit and your DO probe.
  3. Baca Instruksi: Familiarize yourself with the entire maintenance procedure before beginning.

4. Operating Procedures (Probe Maintenance)

Regular maintenance is crucial for accurate DO measurements. Follow these steps for proper probe care:

4.1. Membersihkan Probe

  1. Jarum Bilas: Gently rinse the probe with distilled water to remove any loose debris.
  2. Terapkan Larutan Pembersih: Apply a small amount of the provided cleaning solution to a soft cloth or directly to the probe's membrane area.
  3. Bersihkan dengan Lembut: Carefully wipe the membrane and probe body. For stubborn deposits, use the abrasive foil provided, but apply minimal pressure to avoid damaging the membrane.
  4. Bilas sampai bersih: Rinse the probe thoroughly with distilled water after cleaning to remove all traces of the cleaning solution.

4.2. Replacing the Membrane Head

The membrane head should be replaced periodically or if damaged.

  1. Unscrew Old Head: Carefully unscrew the old membrane head from the probe body.
  2. Siapkan Kepala Baru: Take one of the new membrane heads from the kit.
  3. Fill with Electrolyte: Fill the new membrane head with the provided electrolyte solution. Ensure there are no air bubbles trapped inside. Gently tap the head to dislodge any bubbles.
  4. Screw On New Head: Carefully screw the new membrane head onto the probe body. Ensure it is finger-tight to prevent leaks, but do not overtighten.

4.3. Electrolyte Replacement

The electrolyte solution should be replaced when replacing the membrane head or if the probe's response becomes sluggish.

  1. Drain Old Electrolyte: If not replacing the membrane head, carefully drain the old electrolyte solution from the existing membrane head.
  2. Refill with Fresh Electrolyte: Fill the membrane head with fresh electrolyte solution from the kit, ensuring no air bubbles are present.

5. Penyimpanan dan penanganan

Proper storage of the kit components and the probe itself ensures their longevity and readiness for use.

  • Store the kit components in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Keep bottles of cleaning and electrolyte solutions tightly sealed when not in use.
  • Always store the DO probe with its protective cap filled with a small amount of electrolyte solution or in a suitable storage solution as recommended by WTW.

6. Pemecahan Masalah Umum

If you encounter issues with your DO probe after maintenance, consider the following:

MasalahKemungkinan PenyebabLarutan
Pembacaan tidak stabilAir bubbles in membrane head, dirty membrane, depleted electrolyte.Tap membrane head to remove bubbles, clean membrane, replace electrolyte.
Waktu respons lambatOld or contaminated electrolyte, damaged membrane.Replace electrolyte and/or membrane head.
Tidak ada bacaanCompletely dry membrane, severe probe damage.Refill with electrolyte and re-condition probe. If still no reading, probe may need replacement.

7. Spesifikasi

Key specifications for the WTW 202706 Heavy-Duty DO Probe Maintenance Kit:

  • Nomor Model: 202706
  • Kesesuaian: Designed for WTW CellOx DO probes
  • Isi Paket: Cleaning solution, abrasive foil, 3 membrane heads, electrolyte solution
  • Dimensi Paket: 8.79x5.49x2.64 inci
  • Berat Paket: 9.44 ons
  • Pabrikan: COLE-PARMER (for distribution)
  • Merek: Apa itu WWT?

8. Garansi dan Dukungan

The WTW 202706 Heavy-Duty DO Probe Maintenance Kit comes with a Garansi 1 tahun sejak tanggal pembelian.

For technical support, warranty claims, or further assistance, please contact your authorized WTW distributor or visit the official WTW website. Ensure you have your product model number (202706) and proof of purchase available when seeking support.

Dokumen Terkait - 202706

Praview WTW-PDRP4620EASD3/4630ESD3 dapat digunakan sebagai pengganti
WTW-PDRP4620EASD3 dan WTW-PDRP4630ESD3 OSDメニュー設定、基本接続、および安全上の注意事項をbisnis yang baik.
Praview Panduan Menu dan Koneksi OSD Kamera WTW
Panduan lengkap untuk menghubungkan kamera pengawasan WTW dan mengkonfigurasi pengaturan menu On-Screen Display (OSD), meliputi mode video, pencahayaan, pengaturan siang/malam, dan banyak lagi. Termasuk diagram koneksi terperinci dan instruksi pengaturan.
Praview WTW 取扱説明書:安全上の注意とメニュー設定
WTW adalah salah satu fitur yang paling penting di dunia.関する詳細な取扱説明書。カメラ設定、ビデオモード、言語選択、露出調整などの機能について解説。
Praview WTW-WSP12-BM / WSP18-BM / WSP21-BM IPカメラ 取扱説明書
nama WTW-WSP12-BM, WTW-WSP18-BM, WTW-WSP21-BM IPカメラの取扱説明書。設置、設定、機能、トラブルシューティング、保証に関する詳細情報。
Praview Panduan Pengguna Kamera IP WTW-EGDRY388GWX/GBX
Manual ini menyediakan petunjuk untuk menyiapkan, mendaftarkan, dan menggunakan Kamera IP WTW-EGDRY388GWX/GBX, termasuk instalasi aplikasi, konfigurasi kamera, deteksi gerakan, perekaman, pemutaran, berbagi, dan pemecahan masalah.
Praview WTW 取扱説明書・設定ガイド
WTW adalah sistem operasi yang sangat berguna. cara untuk mendapatkan uang kembalian本語、英語、スペイン語、フランス語に対応。