AEMC 6255

AEMC INSTRUMENTS Model 6255 Micro-Ohmmeter User Manual

Precision Resistance Measurement Device

1. Lebihview

The AEMC INSTRUMENTS Model 6255 is a high-precision Micro-Ohmmeter designed for accurate resistance measurements. It is capable of measuring resistance from 0.1 microhm up to 2500 ohms, with a resolution of 0.1 microhm. This device is suitable for a wide range of industrial and scientific applications requiring precise low-resistance measurements.

Key features include an LCD display, an output current range of 0.1A to 10A, and internal data storage for up to 1500 test results. It is powered by an 8.5Ah NiMH battery, ensuring portability and extended operation.

2. Informasi Keselamatan

Before operating the AEMC Model 6255, please read and understand all safety instructions. Failure to comply with these instructions may result in electric shock, fire, or damage to the instrument.

3. Komponen dan Fitur Produk

The AEMC Model 6255 Micro-Ohmmeter comes with the following items:

AEMC INSTRUMENTS Model 6255 Micro-Ohmmeter in its yellow protective case, open to show the control panel and LCD display.

Figure 1: AEMC INSTRUMENTS Model 6255 Micro-Ohmmeter.The image displays the AEMC Model 6255 Micro-Ohmmeter housed within its robust yellow protective case, which is shown open. The device itself is yellow and black, featuring a clear LCD display that shows "72.6" and "0.1365 mΩ". To the left of the display are red and black input terminals labeled C1, P1, P2, and C2. Below the display are several control buttons. On the right side, there is a large rotary dial and a yellow START/STOP button. The overall design emphasizes durability and ease of use for industrial applications.

Komponen Utama:

4. Pengaturan

4.1 Pengisian Awal

The AEMC Model 6255 is powered by an 8.5Ah NiMH battery. Before first use, or if the battery is low, connect the provided charger to the instrument and a suitable power outlet. Allow sufficient time for the battery to fully charge. The charging indicator on the device will provide status.

4.2 Menghubungkan Kabel Uji

The Model 6255 utilizes a 4-wire (Kelvin) measurement method for high accuracy, which requires four connections to the test object. This method minimizes the impact of lead resistance on the measurement.

  1. Ensure the instrument is turned off before connecting leads.
  2. Connect the current leads (C1 and C2) to the outer points of the resistance to be measured.
  3. Connect the potential leads (P1 and P2) to the inner points of the resistance to be measured. These leads measure the voltage drop across the resistance.
  4. Ensure all connections are secure and clean to prevent erroneous readings.

5. Petunjuk Pengoperasian

5.1 Menghidupkan/Mematikan

Press and hold the power button (if separate, or a specific function on the rotary dial) to turn the instrument on. Repeat the action to turn it off.

5.2 Memilih Mode Pengukuran

Use the rotary selector switch to choose the desired measurement range or mode. The Model 6255 supports resistance measurements from 0.1 microhm to 2500 ohms.

5.3 Melakukan Pengukuran

  1. Ensure the test leads are correctly connected to the object under test as described in Section 4.2.
  2. Select the appropriate measurement range using the rotary dial. If unsure, start with a higher range and decrease as needed.
  3. Tekan tombol MULAI/BERHENTI button to initiate the measurement. The instrument will apply a test current (0.1A to 10A) and display the resistance value on the LCD.
  4. Wait for the reading to stabilize. The LCD will show the measured resistance with 0.1 microhm resolution.
  5. Tekan tombol MULAI/BERHENTI button again to stop the measurement.

5.4 Penyimpanan dan Pengambilan Data

The Model 6255 can store up to 1500 test results. Refer to the on-screen menu or specific buttons (e.g., "MEM" or "SAVE") for instructions on saving and recalling measurements. The RS232 interface allows for data transfer to a computer for further analysis.

6. Pemeliharaan

6.1 Pembersihan

Regularly clean the instrument's casing and display with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the instrument is turned off and disconnected from any power source before cleaning.

6.2 Perawatan Baterai

To prolong the life of the 8.5Ah NiMH battery:

6.3 Kalibrasi

The AEMC Model 6255 is a precision instrument. While the product description states "No Calibration Certificate" is included, periodic calibration by a qualified service center is recommended to ensure continued accuracy and compliance with industry standards. Contact AEMC INSTRUMENTS for calibration services.

7. Penyelesaian masalah

MasalahKemungkinan PenyebabLarutan
Instrumen tidak menyala.Baterai habis.Connect the charger and allow the battery to charge fully.
Erratic or unstable readings.Poor test lead connections; Interference; Test object issues.Ensure leads are securely connected and clean. Move away from strong electromagnetic fields. Verify the integrity of the test object.
Display shows "OVERLOAD" or "OL".Resistance value exceeds selected range; Open circuit.Select a higher measurement range. Check for open circuits in the test setup.
Hasil tes tidak dapat disimpan.Memori penuh.Transfer stored data to a computer via RS232 and clear instrument memory.

8. Spesifikasi

ParameterNilai
Model6255
Rentang Resistensi0.1 microhm to 2500 ohm
Resolusi0.1 microhm
Ketepatan+/-0.05 Percent
Arus Keluaran0.1A hingga 10A
MenampilkanLayar LCD
Sumber Daya8.5Ah NiMH Battery
Hitungan2,500
Peringkat KeamananIEC/EN 61010-1 CAT III 50V Pol 2
Penyimpanan Data1500 Hasil Uji
AntarmukaRS232
Dimensi Produk1x1x1 inci
Berat Barang0.01 Ons
PabrikanINSTRUMEN AEMC

9. Informasi Garansi

AEMC INSTRUMENTS products are designed and manufactured to the highest quality standards. For specific warranty terms and conditions applicable to your Model 6255 Micro-Ohmmeter, please refer to the warranty card included with your product or visit the official AEMC INSTRUMENTS website. Typically, warranties cover defects in materials and workmanship under normal use.

Harap simpan bukti pembelian Anda untuk klaim garansi apa pun.

10. Dukungan Pelanggan

For technical assistance, service, or repair of your AEMC Model 6255 Micro-Ohmmeter, please contact AEMC INSTRUMENTS customer support.

When contacting support, please have your instrument's model number (6255) and serial number ready.

Dokumen Terkait - 6255

Praview AEMC Digital FlexProbe 400D/4000D Series User Manual - AC Current Measurement
User manual for AEMC Digital FlexProbe models 400D-6, 400D-10, 4000D-14, and 4000D-24. Features include TRMS AC current measurement, CAT IV rating, flexible sensor operation, and detailed specifications for electrical, environmental, and mechanical parameters.
Praview AEMC Digital FlexProbe User Manual: Models 400D-6, 400D-10, 4000D-14, 4000D-24
Comprehensive user manual for AEMC Digital FlexProbe series AC current meters (Models 400D-6, 400D-10, 4000D-14, 4000D-24), detailing product features, operation, specifications, safety precautions, and maintenance procedures.
Praview AEMC PEL 52 Power & Energy Logger User Manual
User manual for the AEMC PEL 52 Power & Energy Logger, detailing its features, specifications, operation, setup, maintenance, and troubleshooting for electrical measurements in single-phase and two-phase distribution networks.
Praview AEMC 2620 Ground Fault/Leakage Detector with AN-1 Artificial Neutral User Manual
User manual for the AEMC Model 2620 Ground Fault/Leakage Detector and the AN-1 Artificial Neutral accessory, detailing specifications, operation, troubleshooting, and maintenance.
Praview Panduan Pengguna AEMC MD305 AC Current Probe
User manual for the AEMC MD305 AC Current Probe, detailing its specifications, operation, safety precautions, maintenance, and ordering information for industrial environments.
Praview AEMC AC Current Probe Model MN05 User Manual
User manual for the AEMC AC Current Probe Model MN05, detailing its specifications, operation, safety precautions, maintenance, and warranty. This compact probe offers dual measurement ranges and is designed for industrial and electrical contracting applications.